凯发·k8国际(中国)首页登录


Patient Monitoring

Frequently Asked Questions

1.No AC power supply is connected, or the battery level is low or the battery is damaged →

1.1. Please confirm if the AC power supply is correctly connected;

1.2. Please confirm if the battery level is low or the battery is damaged

2.Connection line fault →

2.1. Check if the connection line between the on/off button board and the main control board is properly connected

2.2. Check if the plug and corresponding socket of the connection line are damaged

3.The on/off button board is damaged → Replace the on/off button board

4.The power module is damaged → Replace the power module

5.The main control board is damaged → Replace the main control board

6.Power protection →

6.1. If the main unit is connected to other devices such as module, external module rack, mouse, keyboard, USB drive, and scanner, please first disconnect these devices from the host. If the main unit can be started up after the disconnection, it can be suspected that the external device fails, triggering power protection of the main unit

6.2. If the main unit is not connected to other devices, check whether there is a short circuit failure with the rear panel of the internal module rack or the main control board, causing protection of the power output

1.The sensor is not installed properly in position or the patient experiences vigorous movement;

2.The ipsilateral arm blood pressure is measured or there is ipsilateral lateral compression;

3.Avoid interference from a strong light environment on signals;

4.Poor peripheral circulation: such as shock and low finger temperature;

5.Fingers: Factors such as nail polish, thick cocoon, broken finger, and long nail affect the light transmittance;

6.Intravenous injection of colored drugs;

7.Determine whether the SpO2 probe is damaged by replacing it

1.Check if the gas circuit is bent or blocked

2.Determine whether the gas transfer tube/cuff is leaking through cross replacement testing

1.Main unit cleaning:

1.1. Dip a dust-free cloth in an appropriate amount of water or 75% ethanol, and squeeze it dry;

1.2. Wipe the display screen;

1.3. Wipe the surface of the main unit, module, or module rack, noting to avoid the equipment interfaces and metal parts;

1.4. Use a dry dust-free cloth to wipe off the cleaning agent from the surface of the equipment, and place the equipment in a cool and ventilated environment to air dry

2.Cleaning of accessories (gas transfer tube for blood pressure, SpO2 cable, SpO2 probe, ECG cable):

2.1. Use a dust-free cloth moistened with water or 75% ethanol to wipe the accessories;

2.2. Use a dry dust-free cloth to wipe off the remaining cleaning agent;

2.3. Place the accessories in a ventilated environment to dry;

2.4. The blood pressure sleeve wrapping cloth can be hand washed using water or mild detergent, and air-dried after cleaning.

3.Main unit disinfection:

3.1. Disinfect the equipment according to the disinfection regulations in the hospital;

3.2. Complete cleaning before disinfection;

3.3. Dilute and use disinfectants according to the manufacturer's instructions:

Ethanol 75%Liquid

Isopropanol 75%Liquid

Sodium hypochlorite 0.5%? ? ? ? ? ? ? ? ? ? ? ?Liquid

Hydrogen peroxide 3%Liquid

N-propanol (1-propanol) 50%Liquid

Quartasept R surface disinfectant Liquid

4.Disinfection of accessories (gas transfer tube for blood pressure, SpO2 cable, SpO2 probe, ECG cable):

PDI Super Sany ClothR disposable sterilization wipe VIRAGUARDR medical surface disinfection wipe

Glutaraldehyde 2% Liquid

Ethanol 75% Liquid

Isopropanol 75% Liquid

N-propanol (1-propanol) 50%Liquid

5.Sleeve disinfection:

You can use a dust-free cloth moistened with 75% ethanol or 75% isopropanol for wiping and disinfection.

A standard quick operation card for the device is provided. For details, refer to BeneVision N17N15N12_Quick Operation Guide_V2.0-CH.

The service life of lithium batteries is three years. Replace the battery after its service life ends. Failing to replace the battery in time may cause the battery to overheat during use of the monitor, leading to serious damage to the device.

1.Check the sticking quality of electrode and reduce patient movement.

2.Adjust the RESP threshold automatically or manually.

Cause analysis:

1.Deviation caused by different measurement principles: ? Mindray NIBP uses the oscillation method for blood pressure measurement. According to the ISO81060-2-2013 non-invasive automated sphygmomanometer standard, the NIBP equipment can be verified by referring to the auscultation or invasive method. ? IBP is a continuous blood pressure measurement method, and its measurement results can reflect the patient's blood pressure changes in real time; NIBP is an intermittent blood pressure measurement method, and its measurement results reflect the blood pressure status over a period of time during the measurement.

2.Deviation caused by different measurement sites: ? A typical measurement method in clinical practice is to take the radial artery of one arm for IBP measurement, and the brachial artery of the other arm for NIBP measurement. Due to differences in the blood pressures of the left and right arms of the human body and different sites of the same arm, the deviations in the IBP and NIBP measurement results can be increased.

3.Deviations caused by non-standard IBP and NIBP measurements: ? IBP sensor incompatibility (zero drift, poor contact) ? Improper piercing position of IBP (the piercing probe tightly adheres to blood vessel wall); ? Improper monitor settings (IBP sensitivity); ? An error found in IBP zero calibration (at the same level as the heart); ? Bubbles or vibrations found in the IBP liquid circuit (affecting accuracy); ? Non-standard NIBP measurement (improper cuff size, external interference)

false